title

ارزیابی عملکرد سیستم برنامه ریزی درمان سه بعدی (TPS) در محاسبه دز میدان های گوه ای مجازی با استفاده از فیلم دزیمتری

زینالی, احد ، گنجگاهی, سمانه ، ملازاده, میکائیل (1396) ارزیابی عملکرد سیستم برنامه ریزی درمان سه بعدی (TPS) در محاسبه دز میدان های گوه ای مجازی با استفاده از فیلم دزیمتری. Iranian Journal of Medical Physics ــ 15 (Specia). ص. 180. شاپا 1735-7241

متن کامل

[img]
پیش نمایش
متنی - نسخه چاپ شده
127kB

آدرس اینترنتی رسمی : https://ijmp.mums.ac.ir/article_12792_1314.html


عنوان انگليسی

Evaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry

خلاصه انگلیسی

Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelerator supports a virtual wedge ، that this wedge creates a dose distribution similar to a physical wedge without the use of any extra accessory and wedged profiles are produced moving collimator jaw during irradiation with a constant speed but varying the dose rate. In this study ، TPS performance is evaluated in virtual wedged fields by comparing the calculated and measured results. Materials and Methods: The calculations were performed by the collapsed cone superposition algorithm based TPS for two tangent fields in anthropomorphic slab phantom using 15،̊ 30̊، 45̊، 60̊ virtual wedges for field size of 20 × 20 cm2 for 6 and 15 MV photon beams. Measurements were produced by Gafchromic EBT films and reading exposed films with scanner. Results: Good agreement between the measured dose with film dosimetry and calculated dose with collapsed cone superposition algorithm based TPS with using virtual wedge in heterogeneous environment and different energies were found ، with the maximum difference not exceed 3 -5%. The increase in wedge angle due to increase in the difference between calculated and measured data. Conclusion: : The results from this study showed that the accuracy of collapsed cone superposition algorithm based TPS used with the virtual wedges for two tangent fields is enough for the clinical usage under the studied experimental conditions.

نوع سند :مقاله
زبان سند : انگلیسی
نویسنده مسئول :احد زینالی
نویسنده :سمانه گنجگاهی
نویسنده :میکائیل ملازاده
ضریب تاثیر و نمایه مجلات:Indexed in: SID, Magiran
کلیدواژه ها (انگلیسی):virtual wedge Treatment planning systems(TPS) Gafchromic film film dosimetry
موضوعات :WN پرتو شناسی
بخش های دانشگاهی :دانشكده پزشكي > گروه علوم پایه > بخش فیزیک پزشکی