رضاپور, جعفر ، مستعار, احمد ، طریقت نیا, علی ، فلاحتی, فرناز ، حسینی, سید محمد ، Johal, Grokan ، Nader, Nader D ، دسترنج, لیلا ، عابدی, رزاق (1400) تهاتر بین میانگین دوز غده پستان و کیفیت تصویر در سیستم های ماموگرافی دیجیتال و معمولی: یک مطالعه چند مرکزی. Radioprotection ــ 56 (3). ص.ص.221-227. شاپا 0033-8451

متن کامل

[img] متنی - نسخه چاپ شده
محدود به فقط پرسنل سامانه

734kB

آدرس اینترنتی رسمی : https://www.researchgate.net/profile/Nader-Nader-2...


عنوان انگليسی

Trade-off between breast mean glandular dose and image quality in digital and conventional mammogram systems: a multicenter study

خلاصه انگلیسی

Evaluating the patient dose or exposure parameters considering the image quality can improve the chances of accurate diagnosis and reduce the unnecessary exposures from medical devices such as mammography. This study aimed to evaluate digital and conventional mammography machines while considering the trade-off between image quality and mean glandular dose (MGD) using a phantom. In the present study, one full-field digital mammography (FFDM) and two film-screen mammography (FSM) machines were investigated. The MGD values and image quality were assessed using the American College of Radiology (ACR) phantom at various mAs and constant kVp values. The results were obtained and compared with European guidelines. Friedman and Wilcoxon statistical tests were used to show the comparison. The results from the quality control (QC) tests demonstrated that all machines are functioning well. The best image quality in the digital mammography machine was observed at the MGD of 1.8 mGy and 55 mAs. In addition, the two conventional machines had the best image quality regarding the imaging of the ACR phantom at 65 mAs with an MGD of 2.1 mGy. These values were considered as appropriate values for the studied mammography systems. Furthermore, the Friedman test demonstrated that there are significant differences between the measured image quality values obtained from the different machines (p < 0.05), however, according to the Wilcoxon test there were not any significant differences between the conventional machines at various mAs values. Owing to the results, for a medium breast size, the image quality will not be improved with increasing the exposure after a specified MGD corresponds to a certain mAs. It is notable that this value is smaller in digital mammography system at a reasonably low dose.

نوع سند :مقاله
زبان سند : انگلیسی
نویسنده اول :جعفر رضاپور
نویسنده :احمد مستعار
نویسنده :علی طریقت نیا
نویسنده :فرناز فلاحتی
نویسنده :سید محمد حسینی
نویسنده :Grokan Johal
نویسنده :Nader D Nader
نویسنده مسئول :لیلا دسترنج
نویسنده مسئول :رزاق عابدی
ضریب تاثیر و نمایه مجلات:IF: 1.1 Indexed in: ISI, Scopus
کلیدواژه ها (فارسی):ماموگرافی دیجیتال، ماموگرافی معمولی،کیفیت تصویر،میانگین دوز غده ای پستان
کلیدواژه ها (انگلیسی):Digital mammography , Conventional mammography , Image quality , Mean glandular dose
موضوعات :WN پرتو شناسی
بخش های دانشگاهی :دانشكده پزشكي > گروه علوم پایه > بخش فیزیک پزشکی
کد شناسایی :17364
ارائه شده توسط : دکتر علی طریقت نیا
ارائه شده در تاریخ :15 مرداد 1402 14:37
آخرین تغییر :15 مرداد 1402 14:37

فقط پرسنل کتابخانه صفحه کنترل اسناد

Document Downloads

More statistics for this item...